Title of article
Soft X-ray charger (SXC) system for use with electrospray for mobility measurement of bioaerosols
Author/Authors
Luis B. Modesto-Lopez، نويسنده , , Eric M. Kettleson، نويسنده , , Pratim Biswas، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2011
Pages
8
From page
357
To page
364
Abstract
Electrospray size spectrometry analysis (ESSA), a mobility-based aerosol technique to characterize the size and charge-state of nanoparticles, uses a neutralizer to bring multiply charged particles to an equilibrium charge-state. Typical ESSA setups incorporate a Po-210 neutralizer that has short half-life time and its radioactive nature leads to safety concerns. This work demonstrates that a soft X-ray charger (SXC) effectively neutralizes electrosprayed bioparticles. Calculations show that the SXC generates bipolar ions with number concentrations one and two orders of magnitude higher than conventional Po-210 and Kr-85 sources, respectively. Additionally, we established that an SXC may be safely implemented in ESSA.
Keywords
Proteins , Photoionization , Sizespectrometryanalysis , electrospray , SoftX-raycharger , viruses
Journal title
JOURNAL OF ELECTROSTATICS
Serial Year
2011
Journal title
JOURNAL OF ELECTROSTATICS
Record number
1265293
Link To Document