• Title of article

    In search of a standard method for the characterisation of organic solvent nanofiltration membranes

  • Author/Authors

    Y.H. See Toh، نويسنده , , X.X. Loh، نويسنده , , K. Li، نويسنده , , A. Bismarck، نويسنده , , A.G. Livingston، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2007
  • Pages
    6
  • From page
    120
  • To page
    125
  • Abstract
    This paper describes a method for the determination of the relationship between solute molecular weight and rejection, often referred to as the molecular weight cut off (MWCO) of the organic solvent nanofiltration (OSN) membranes. This method utilises a homologous series of styrene oligomers spanning the nanofiltration range (200–1000 g mol−1). Good separation of the individual oligomers was achieved using liquid chromatography with a reverse phase C18 column. The good solubility of the styrene oligomers in many organic solvents makes cross comparison of membrane performance across these solvents possible. This method also allows the determination of the molecular weight at which a rejection of >99.9% occurs in different solvents, making it a useful tool to test membrane and equipment integrity.
  • Keywords
    Organic solvent nanofiltration , MWCO , Characterisation , Starmem , Styrene oligomers
  • Journal title
    Journal of Membrane Science
  • Serial Year
    2007
  • Journal title
    Journal of Membrane Science
  • Record number

    1353039