Title of article :
Modification of the ELPI to measure mean particle effective density in real-time
Author/Authors :
Rostedt، نويسنده , , Antti and Marjamنki، نويسنده , , Marko and Keskinen، نويسنده , , Jorma، نويسنده ,
Issue Information :
ماهنامه با شماره پیاپی سال 2009
Pages :
9
From page :
823
To page :
831
Abstract :
A new modification of electrical low pressure impactor (ELPI) for the particle effective density measurement is presented. The system is capable of real-time operation and it is based on the serial measurement of mobility and aerodynamic diameter. In the studied configuration, a zeroth order mobility analyser is installed inside of the ELPI-instrument. The system is feasible for single modal distributions. For several particle materials and varying size distributions, the measured average density values were within 15% of the values obtained with a reference method.
Keywords :
impactor , Density , Mobility
Journal title :
Journal of Aerosol Science
Serial Year :
2009
Journal title :
Journal of Aerosol Science
Record number :
1385443
Link To Document :
بازگشت