Title of article :
Development and performance test of a micromachined unipolar charger for measurements of submicron aerosol particles having a log-normal size distribution
Author/Authors :
Park، نويسنده , , Dongho and Kim، نويسنده , , Yong-Ho and Lee، نويسنده , , Sang-Goo and Kim، نويسنده , , Chul-Gyun Hwang، نويسنده , , Jungho and Kim، نويسنده , , Yong-Jun، نويسنده ,
Issue Information :
ماهنامه با شماره پیاپی سال 2010
Abstract :
We designed and evaluated a chip-type unipolar charger based on a micro-electro-mechanical systems (MEMS) process for a compact, portable, inexpensive, and easy-to-deploy instrument to measure real-time aerosol particle size distribution with reasonable accuracy. The charger was evaluated using two parameters: particle loss and charging characteristics which were the product of penetration (p) and average number of charges (np). Particle losses were below 16.6%, and p·np was approximately proportional to d p 2 . After evaluation, the number concentrations and the geometric mean diameters of NaCl aerosol particles were estimated using a method presented in Park, An, and Hwang (2007a) and were compared to scanning mobility particle sizer (SMPS) data. To simulate the charger, a commercial computational fluid dynamics (CFD) software, FLUENT (6.3 version), with external user defined function (UDF) code, was used to solve the electric field, flow field, average particle charge, and particle trajectories. The measured data of particle loss, average particle charge, number concentration, and geometric mean diameter were in good agreement with the results calculated by FLUENT.
Keywords :
Miniature device , micro-electro-mechanical systems , Corona discharge , Unipolar charging , Particle size distribution
Journal title :
Journal of Aerosol Science
Journal title :
Journal of Aerosol Science