• Title of article

    Towards traceable particle number concentration standard: Single charged aerosol reference (SCAR)

  • Author/Authors

    Yli-Ojanperن، نويسنده , , Jaakko and Mنkelن، نويسنده , , Jyrki M. and Marjamنki، نويسنده , , Marko and Rostedt، نويسنده , , Antti and Keskinen، نويسنده , , Jorma، نويسنده ,

  • Issue Information
    ماهنامه با شماره پیاپی سال 2010
  • Pages
    10
  • From page
    719
  • To page
    728
  • Abstract
    A concept of realizing a standard for aerosol particle number concentration was tested, based on generating singly charged aerosol particles in the size range from 10 up to 500 nm. To this end, a device named single-charged aerosol reference (SCAR) was designed, built, and tested. The device is based on electrical charging of nanoparticles and subsequent growth of the particles. With an accurate measurement of volume flow and electrical current from the singly charged particles, the number concentration can be accurately, and in the end, traceably determined. Laboratory tests have shown that the device can be used to generate a narrow (GSD<1.3) particle size distribution of singly charged particles. The device can be used for traceable calibration of instruments measuring the number concentration of the particles.
  • Keywords
    Singly charged , Aerosol reference , Standard , Number concentration , Calibration
  • Journal title
    Journal of Aerosol Science
  • Serial Year
    2010
  • Journal title
    Journal of Aerosol Science
  • Record number