    Precision of PM measurements with the UNC passive aerosol sampler

  • Author/Authors

    Arashiro، نويسنده , , Maiko and Leith، نويسنده , , David، نويسنده ,

  • Issue Information
    ماهنامه با شماره پیاپی سال 2013
  • Pages
    4
  • From page
    181
  • To page
    184
  • Abstract
    The UNC passive aerosol sampler measures concentrations and size distributions of indoor, outdoor, and occupational particulate matter. This study assessed the precision of passive sampler measurements using data from duplicate pairs of samplers that were deployed in a large field study. Particle count and size data were used to calculate PM2.5 and PM10-2.5 concentrations. To determine precision, the concentrations measured by duplicate samplers were compared to one another. Results showed that this sampler provides mass concentrations of PM2.5 and PM10-2.5 in the field with relative standard deviations that approach about 15%, with greater precision associated with higher concentrations (PM2.5>5 μg/m3 and PM10-2.5>20 μg/m3). Variability between measurements was associated with counting statistics and possibly with differences in the open area of the passive sampler mesh screen.
  • Keywords
    Measurement precision , aerosols , UNC passive aerosol sampler , Particulate matter