Title of article :
Virtual evaluation of stent graft deployment: A validated modeling and simulation study
Author/Authors :
De Bock، نويسنده , , S. and Iannaccone، نويسنده , , F. and De Santis، نويسنده , , G. and De Beule، نويسنده , , M. and Van Loo، نويسنده , , D. and Devos، نويسنده , , D. and Vermassen، نويسنده , , F. and Segers، نويسنده , , P. and Verhegghe، نويسنده , , B.، نويسنده ,
Issue Information :
ماهنامه با شماره پیاپی سال 2012
Abstract :
The presented study details the virtual deployment of a bifurcated stent graft (Medtronic Talent) in an Abdominal Aortic Aneurysm model, using the finite element method. The entire deployment procedure is modeled, with the stent graft being crimped and bent according to the vessel geometry, and subsequently released. The finite element results are validated in vitro with placement of the device in a silicone mock aneurysm, using high resolution CT scans to evaluate the result. The presented work confirms the capability of finite element computer simulations to predict the deformed configuration after endovascular aneurysm repair (EVAR). These simulations can be used to quantify mechanical parameters, such as neck dilations, radial forces and stresses in the device, that are difficult or impossible to obtain from medical imaging.
Keywords :
Stent graft , Silicone validation , deployment , Finite element , Abdominal Aortic Aneurysm
Journal title :
Journal of the Mechanical Behavior of Biomedical Materials
Journal title :
Journal of the Mechanical Behavior of Biomedical Materials