Title of article :
An amperometric silicon-based biosensor for d-lactate
Author/Authors :
Tap، نويسنده , , H and Gros، نويسنده , , P and Gué، نويسنده , , A.M، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2000
Pages :
5
From page :
123
To page :
127
Abstract :
A miniaturised biosensor for d-lactate has been fabricated on silicon wafer. Microsystems technologies led to enhanced efficiency in terms of sample size and reagent consumption while equally allowing mass production. The biosensor is based on the association of an enzymatic solution including diaphorase and d-lactate dehydrogenase confined in the microstructure by a membrane and two weakly polarised gold microelectrodes. Various electrochemical pretreatments of the gold surface are performed to increase the deviceʹs sensitivity. Depending on the geometry involved, the proposed biosensor features a response time ranging from 2 to 5 min, a sensitivity of 2 or 8 μA mM−1 cm−2 and a linearity range from 5×10−3 to 1.5 mmol l−1. Its lifetime is 3 weeks with 10 assays/day.
Keywords :
Amperometric biosensor , Microsystems technologies , Enzymatic catalysis , d-lactate assay