Title of article :
An integrated CMOS microluminometer for low-level luminescence sensing in the bioluminescent bioreporter integrated circuit
Author/Authors :
Simpson، نويسنده , , Michael L and Sayler، نويسنده , , Gary S and Patterson، نويسنده , , Greg and Nivens، نويسنده , , David E and Bolton، نويسنده , , Eric K and Rochelle، نويسنده , , James M and Arnott، نويسنده , , James C and Applegate، نويسنده , , Bruce M and Ripp، نويسنده , , Steven and Guillorn، نويسنده , , Michael A، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2001
Pages :
7
From page :
134
To page :
140
Abstract :
We report an integrated CMOS microluminometer for the detection of low-level bioluminescence in whole cell biosensing applications. This microluminometer is the microelectronic portion of the bioluminescent bioreporter integrated circuit (BBIC). This device uses the n-well/p-substrate junction of a standard bulk CMOS IC process to form the integrated photodetector. This photodetector uses a distributed electrode configuration that minimizes detector noise. Signal processing is accomplished with a current-to-frequency converter circuit that forms the causal portion of the matched filter for dc luminescence in wide-band white noise. Measurements show that luminescence can be detected from as few as 4×105 cells/ml.
Keywords :
Biosensors , Luminometry , Whole-cell biosensors , CMOS microluminometer , Bioluminescence
Journal title :
Sensors and Actuators B: Chemical
Serial Year :
2001
Journal title :
Sensors and Actuators B: Chemical
Record number :
1412690
Link To Document :
بازگشت