Title of article :
Temperature analysis of continuous-flow micro-PCR based on FEA
Author/Authors :
Zhang، نويسنده , , Qintao and Wang، نويسنده , , Wenhui and Zhang، نويسنده , , Hongshen and Wang، نويسنده , , Yuelin، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2002
Abstract :
Continuous-flow micro-PCR chip is a novel method, which is quite different from conventional PCR and micro-chamber PCR chip. A fast and continuous DNA amplification reaction can be obtained by moving the sample through three individual temperature zones whose temperature is constant during the course of reaction. In this paper, temperature patterns of two kinds of continuous-flow micro-PCR: glass–glass-bonding chip and silicon–glass-bonding chip are studied using finite element analysis (FEA). The temperature uniformity and temperature gradient on the chips surface are investigated. Based on analysis results, a simple method that can realize continuous-flow PCR on silicon–glass-bonding chip is introduced. Only denaturation zone is heated and the water is flowed through the renaturation zone and extension zone along a channel. Preliminary experiment is designed to prove the feasibility of the method. And an interface chip is designed to realize multi-times PCR using one chip.
Keywords :
Finite element analysis (FEA) , Polymerase chain reaction (PCR) , Micro-electro-mechanical system (MEMS) , Micro-PCR
Journal title :
Sensors and Actuators B: Chemical
Journal title :
Sensors and Actuators B: Chemical