Title of article :
Thick film sensors based on laccases from different sources immobilized in polyaniline matrix
Author/Authors :
Timur، نويسنده , , Suna and Pazarl?o?lu، نويسنده , , Nurdan and Pilloton، نويسنده , , Roberto and Telefoncu، نويسنده , , Azmi، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2004
Pages :
5
From page :
132
To page :
136
Abstract :
Thick film electrode based biosensors containing Trametes versicolor (TvL) and Aspergillus niger (AnL) laccases and Agaricus bisporus tissues (AbT) were developed for the determination of phenolic compounds and the measurement was based on oxygen consumption in relation to analyte oxidation. The electrodeposited organic polymer; polyaniline was used as a matrix for the immobilization in the preparation of thick film sensors. The systems were calibrated for different phenolic substances. A linearity was obtained in concentration range between 0.4 and 6.0 μM phenol, 0.2 and 1.0 μM catechol, 2.0 and 20.0 μM l-DOPA for TvL based biosensor; for AnL based enzyme electrode 0.4 and 4.0 μM phenol, 0.4 and 15 μM catechol, 0.4 and 6.0 μM l-DOPA; for AbT electrode 1.0 and 10 μM phenol, 0.4 and 1.6 μM catechol, 1.0 and 10 μM l-DOPA, respectively, in the response time of 300 s. Furthermore, as well as sample application and accuracy, optimum pH, temperature and thermal stabilities of the proposed systems were also detected.
Keywords :
Thick film electrodes , Polyaniline film , Phenolic compounds , Laccase
Journal title :
Sensors and Actuators B: Chemical
Serial Year :
2004
Journal title :
Sensors and Actuators B: Chemical
Record number :
1413321