Title of article :
Amperometric biosensor for formic acid in air
Author/Authors :
Sandstrِm، نويسنده , , K.J.Mattias and Newman، نويسنده , , Jeffrey and Sunesson، نويسنده , , Anna-Lena and Levin، نويسنده , , Jan-Olof and Turner، نويسنده , , Anthony P.F. Turner، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2000
Abstract :
The possibility of developing a simple, inexpensive and specific personal passive “real-time” air sampler incorporating a biosensor for formic acid was investigated. The sensor is based on the enzymatic reaction between formic acid and formate dehydrogenase (FDH) with nicotinamide adenine dinucleotide (NAD+) as a co-factor and Meldolaʹs blue as mediator. An effective way to immobilise the enzyme, co-factor and Meldolaʹs blue on screen-printed, disposable, electrodes was found to be in a mixture of glycerol and phosphate buffer covered with a gas-permeable membrane. Steady-state current was reached after 4–15 min and the limit of detection was calculated to be below 1 mg/m3. However, the response decreased by 50% after storage at −15°C for 1 day.
Keywords :
Gas-permeable membrane , Glycerol , Air monitoring , Biosensor , Formic acid
Journal title :
Sensors and Actuators B: Chemical
Journal title :
Sensors and Actuators B: Chemical