• Title of article

    An exchangeable Si wafer internal reflection element for FT-IR measurement of water sample

  • Author/Authors

    Seyama، نويسنده , , Michiko and Sugimoto، نويسنده , , Iwao and Katoh، نويسنده , , Tadashi، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2001
  • Pages
    6
  • From page
    54
  • To page
    59
  • Abstract
    A lipophilic-layer-coated Si internal reflection element (SRE) is reported as an exchangeable probe for Fourier-transform infrared spectroscopic measurement of organic contaminants in water. The SRE is installed in the reflection cell by being stacked on a light-guiding crystal. The lipophilic layer is a self-assembled layer made from octadecyltrichlorosilane. We theoretically and experimentally confirmed that the cell comprising the SRE placed on a ZnSe-light-guiding crystal could measure n-hexane dispersed in water.
  • Keywords
    Si wafer , Fourier-transform infrared spectroscopy , Internal reflection , Lipophilic layer , water
  • Journal title
    Sensors and Actuators B: Chemical
  • Serial Year
    2001
  • Journal title
    Sensors and Actuators B: Chemical
  • Record number