• Title of article

    OxiCarbo®, a single sensor for the non-invasive measurement of arterial oxygen saturation and CO2 partial pressure at the ear lobe

  • Author/Authors

    T. Gisiger، نويسنده , , P.A and Palma، نويسنده , , J.P and Eberhard، نويسنده , , P، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2001
  • Pages
    4
  • From page
    527
  • To page
    530
  • Abstract
    The paper describes the characteristics and results obtained in adults with a new miniaturised sensor combining the measurement of arterial oxygen saturation by optical means and arterial PCO2 by electrochemical means (OxiCarbo® sensor). The sensor is heated to 42°C to create arterialisation of the cutaneous tissue and is applied to the ear lobe with a special low-pressure clip. The use of the sensor heating significantly improves the performance of the SpO2 measurement. The heating increases the arterial pulsation signal and decreases the non-arterialised tissue signal. The fluctuations of the measurement are reduced. The quantitative results of the SpO2 measurements are not affected by the heating. asurement at the ear lobe improves also the PCO2 measurement by decreasing the arterialisation time from approximately 7 min at the forearm to less than 3 min at the ear lobe. At a temperature of 42°C, the sensor may be left at the same central site for over 12 h without the risk of a skin burn.
  • Keywords
    Continuous non-invasive monitoring , oxygen saturation , Carbon dioxide tension , Measurement techniques: pulse oximetry , Transcutaneous PCO2 measurement
  • Journal title
    Sensors and Actuators B: Chemical
  • Serial Year
    2001
  • Journal title
    Sensors and Actuators B: Chemical
  • Record number