• Title of article

    Method for lifetime-based chemical sensing using the demodulation of the luminescence signal

  • Author/Authors

    Andrzejewski، نويسنده , , Damian and Klimant، نويسنده , , Ingo and Podbielska، نويسنده , , Halina، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2002
  • Pages
    7
  • From page
    160
  • To page
    166
  • Abstract
    A novel method for measuring the phosphorescence decay time of optical sensors is presented. It is based on the frequency domain measurement of the demodulation of luminescence signals. Two lock-in amplifiers are used in the set-up. The ratio of two amplitudes simultaneously measured at different modulation frequencies (f1 and f2) corresponds to the luminescence demodulation and therefore to the decay time. asibility of the new method is demonstrated for an optical oxygen sensor. The decay time data obtained from the demodulation measurement are compared with those derived from the conventional phase detection. The well established two-site model for description of intensity calibration curves for oxygen partial pressure, was found to be also suitable to describe the decay time based calibration curves derived from the demodulation and from phase angles.
  • Keywords
    Frequency domain demodulation , Decay time sensing , Oxygen optode
  • Journal title
    Sensors and Actuators B: Chemical
  • Serial Year
    2002
  • Journal title
    Sensors and Actuators B: Chemical
  • Record number

    1417000