Title of article :
Photocurable membranes for ion-selective light-addressable potentiometric sensor
Author/Authors :
Ermolenko، نويسنده , , Yu. and Yoshinobu، نويسنده , , T. and Mourzina، نويسنده , , Yu. and Levichev، نويسنده , , S. and Furuichi، نويسنده , , K. and Vlasov، نويسنده , , Yu and Schِning، نويسنده , , M.J. and Iwasaki، نويسنده , , H.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2002
Pages :
7
From page :
79
To page :
85
Abstract :
A new type of light-addressable potentiometric sensor (LAPS) has been developed by elaboration of photocurable ion-sensitive membranes based on matrices of mono- and diacrylates in combination with a silicon-based electrochemical transducer for the purposes of chemical and biochemical analyses. The characteristics of the elaborated K-, Li-, Cs-, Ca- and Mg-LAPS are presented to indicate the potential of the technology developed. For example, the sensitivities of the K- and Li-LAPS were found to be 58±1 mV/pK and 59±1 mV/pLi, respectively, over concentration ranges of 10−5–10−1 mol/l. The detection limits were evaluated to be 2×10−6 mol/l of the primary ion.
Keywords :
Light-addressable potentiometric sensor (LAPS) , Photocurable membranes , Ion-sensitivity