• Title of article

    Quantification of binary mixtures of the freones R22 and R134a by surface plasmon resonance

  • Author/Authors

    Busche، نويسنده , , Stefan and Dieterle، نويسنده , , Frank and Kieser، نويسنده , , Birgit and Gauglitz، نويسنده , , Günter، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2003
  • Pages
    7
  • From page
    192
  • To page
    198
  • Abstract
    For the recycling process of freones a high purity of the analytes has to be guaranteed. A fast and inexpensive online monitoring is desirable. In this study the quantification of binary mixtures of the freones R22 and R134a by surface plasmon resonance (SPR) was studied using a self-built Kretschmann set-up. A microporous polycarbonate was used as sensitive layer. Due to a size dependent sorption and desorption kinetics of the two freones a quantitative characterization of binary mixtures is possible by the use of only one sensitive layer. 441 different binary mixtures for calibration and 400 test mixtures were measured over time. The measured signals were subdivided into points of time and used as input variables for the data analysis by neural networks. An optimization of the number and type of these points of time resulted in a small relative RMSE of prediction of 1.68% for R22 and 2.33% for R134a for the test mixtures.
  • Keywords
    surface plasmon resonance (SPR) , polycarbonate , time-resolved measurement , NEURAL NETWORKS , Multicomponent analysis , Freones
  • Journal title
    Sensors and Actuators B: Chemical
  • Serial Year
    2003
  • Journal title
    Sensors and Actuators B: Chemical
  • Record number

    1417184