Title of article :
Hydrogen-selective microelectrodes based on silicon needles
Author/Authors :
Zine، نويسنده , , N. and Bausells، نويسنده , , J. and Ivorra، نويسنده , , A. and Aguilَ، نويسنده , , J. and Zabala، نويسنده , , M. and Teixidor، نويسنده , , F. and Masalles، نويسنده , , C. and Viٌas، نويسنده , , C. and Errachid، نويسنده , , A.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2003
Pages :
7
From page :
76
To page :
82
Abstract :
The fabrication of hydrogen-selective microelectrodes on silicon needle-shaped substrates is described. The microelectrodes are based on an ion-selective poly(vinyl chloride) (PVC) membrane with an intrinsically conducting polymer (polypyrrole (PPy)) solid contact layer. The polypyrrole is prepared with the dopant anion cobaltbis(dicarbollide) [3,3′-Co(1,2-C2B9H11)2]−, which gives a high stability to the polymer layer. The performance of the resulting solid-contact ion-selective microelectrodes (SCISME) is investigated by using potentiometric measurement and electrochemical impedance spectrometry. The feasibility of the fabrication technology is demonstrated and the devices operate satisfactorily, with a response showing good sensitivity and selectivity against common interfering cations in background solutions. The SCISME has been developed for organ monitoring during cardiac surgery or during transportation for transplants.
Keywords :
Solid-state ion-selective microelectrode , Conducting organic polymer , Polypyrrole , Silicon needle , pH microelectrode