• Title of article

    Coated wire lead(II) selective potentiometric sensor based on 4-tert-butylcalix[6]arene

  • Author/Authors

    Bhat، نويسنده , , Vaishali S. and Ijeri، نويسنده , , Vijaykumar S. and Srivastava، نويسنده , , Ashwini K.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2004
  • Pages
    8
  • From page
    98
  • To page
    105
  • Abstract
    A coated-wire lead ion-selective electrode (ISE) has been constructed using 5,11,17,23,29,35-hexa-tert-butyl-37,38,39,40,41,42-hexahydroxycalix[6]arene as neutral carrier in a poly(vinyl chloride) matrix. The sensor exhibited a linear Nernstian response over the range 1×10−6 to 1×10−1 M with a slope of 30±1 mV per decade change and a detection limit of 6×10−7 M. The working pH range of the sensor is 1.5–6.0 and it shows good selectivity for Pb(II) ions over other mono-, bi- and trivalent cations. The selectivity coefficients (Kij) were determined for Na(I), K(I), Ag(I), Mg(II), Ca(II), Ba(II), Sr(II), Mn(II), Co(II), Ni(II), Cu(II), Zn(II), Cd(II), Hg(II), Cr(III), Fe(III) and Al(III); the values for these ions being in the range of 10−4–10−2. The sensor was successfully used as an indicator electrode in potentiometric titration of lead ions with EDTA. The ISE was also used for determination of lead in electroplating bath solutions, alloy sample, battery waste samples and effluent waters.
  • Keywords
    41 , Potentiometry , Coated wire ion-selective electrode , 29 , 35-Hexa-tert-butyl-37 , Lead , 17 , 5 , 40 , 39 , 11 , 38 , 23
  • Journal title
    Sensors and Actuators B: Chemical
  • Serial Year
    2004
  • Journal title
    Sensors and Actuators B: Chemical
  • Record number