• Title of article

    The influence of AFM and VSI techniques on the accurate calculation of tribological surface roughness parameters

  • Author/Authors

    Spencer، نويسنده , , A. and Dobryden، نويسنده , , I. and Almqvist، نويسنده , , N. and Almqvist، نويسنده , , A. and Larsson، نويسنده , , R.، نويسنده ,

  • Issue Information
    ماهنامه با شماره پیاپی سال 2013
  • Pages
    9
  • From page
    242
  • To page
    250
  • Abstract
    Vertical Scanning Interferometry (VSI) may induce optical artefacts in surface topography measurements. The influence of these optical artefacts on the calculation of Rk surface roughness parameters, contact stiffness and flow factors were studied. Two surface measurement techniques were used: Atomic Force Microscopy (AFM) and VSI. Calibration grids were used to make it easier to isolate the causes of these artefacts, while a real engineering surface was used to compare these two techniques in an industrially applied case. It was found that the optical artefacts have a large influence on all the roughness parameters, contact stiffness and flow factors calculated on the calibration grids. However, for the engineering surface the differences between AFM and VSI measurements were much smaller.
  • Keywords
    Surface roughness , AFM , VSI , Flow Factors
  • Journal title
    Tribology International
  • Serial Year
    2013
  • Journal title
    Tribology International
  • Record number