    Computational development of a polyethylene wear model for the articular and backside surfaces in modular total knee replacements

  • Author/Authors

    OBrien، نويسنده , , S. and Luo، نويسنده , , Y. and Wu، نويسنده , , C. and Petrak، نويسنده , , M. and Bohm، نويسنده , , E. and Brandt، نويسنده , , J.-M.، نويسنده ,

  • Issue Information
    ماهنامه با شماره پیاپی سال 2013
  • Pages
    8
  • From page
    284
  • To page
    291
  • Abstract
    A computational model to predict polyethylene wear in modular total knee replacements was developed. The results from knee simulator wear tests were implemented with finite element simulations to identify the wear factors of Archardʹs wear law. The calculated wear factor for the articular and backside surface was 1.03±0.22×10−7 mm3/Nm and 2.43±0.52×10−10 mm3/Nm, respectively. The difference in wear factors was attributed to differences in wear mode and wear mechanisms between the articular (mainly two-body rolling/sliding wear mode with an abrasive/adhesive wear mechanism) and the backside surfaces (mainly fretting wear mode with an adhesive wear mechanism).
  • Keywords
    Polyethylene , WEAR , Finite element , total knee replacement