Title of article :
Biotribocorrosion: Some electrochemical observations from an instrumented hip joint simulator
Author/Authors :
Hesketh، نويسنده , , James C Hu، نويسنده , , Xinming and Yan، نويسنده , , Yu and Dowson، نويسنده , , Duncan and Neville، نويسنده , , Anne، نويسنده ,
Issue Information :
ماهنامه با شماره پیاپی سال 2013
Pages :
7
From page :
332
To page :
338
Abstract :
36 mm Metal-on-Metal heads and cups were tested in a hip joint simulator instrumented with a 3-electrode electrochemical cell. This facilitated in situ monitoring of the tribological depassivation of the bearing surface. Subsurface analysis of the different regions of the femoral head was analysed using transmission electron microscopy. It was found that tribological depassivation increased the rate of corrosion, but subsequently decreased at around 450,000 cycles. Observations of the subsurface microstructure revealed the formation of nano-crystallites and the presence of micro-scale subsurface cracks.
Keywords :
Tribocorrosion , Depassivation , Simulator , Hip
Journal title :
Tribology International
Serial Year :
2013
Journal title :
Tribology International
Record number :
1426859
Link To Document :
بازگشت