Title of article :
Calibration of linear encoders with sub-nanometer uncertainty using an optical-zooming laser interferometer
Author/Authors :
Kajima، نويسنده , , Mariko and Minoshima، نويسنده , , Kaoru، نويسنده ,
Issue Information :
فصلنامه با شماره پیاپی سال 2014
Pages :
6
From page :
769
To page :
774
Abstract :
The nonlinear errors of high-precision linear encoders were calibrated by using a nanometer-length calibrator that was based on the optical-zooming laser interferometer with an optical frequency comb. A transmission-type linear encoder and a reflection-type linear encoder were calibrated, and the cyclic nonlinear errors were evident. The magnitudes of the observed cyclic errors were 0.1 nm and 0.2 nm, respectively, and the best calibration uncertainties were 0.55 nm (k = 2). A traceable calibration service for linear encoders with the best calibration uncertainty in the sub-nanometer range has started based on this work.
Keywords :
Length metrology , linear encoder , laser interferometer , Optical frequency comb , Optical zooming
Journal title :
Precision Engineering
Serial Year :
2014
Journal title :
Precision Engineering
Record number :
1430000
Link To Document :
بازگشت