Author/Authors :
Wagner، نويسنده , , Torsten and Maris، نويسنده , , Rob J. and Ackermann، نويسنده , , Hans-Josef and Otto، نويسنده , , Ralph and Beging، نويسنده , , Stefan and Poghossian، نويسنده , , Arshak and Schِning، نويسنده , , Michael J.، نويسنده ,
Abstract :
A user-friendly handheld measurement device in combination with a field-effect-based capacitive electrolyte–insulator–semiconductor (EIS) sensor has been developed. The measurement device allows the characterisation of various chemical sensors and biosensors in both the capacitance–voltage (C/V) and the constant-capacitance (ConCap) mode. The developed device has been systematically tested with pH-sensitive EIS sensors in terms of accuracy, reliability and noise behaviour. Many possible applications for this type of semiconductor-based sensor with the associated convenient handheld measurement device can be found in the biotechnology environment. First practical aspects and considerations to apply EIS-based sensors together with the new handheld measurement device for pH measurements in a cell-culture medium will be reported.