• Title of article

    Different calibration methods of a microbial BOD sensor for analysis of municipal wastewaters

  • Author/Authors

    Velling، Pirkko نويسنده , , Siiri and Tenno، نويسنده , , Toomas، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2009
  • Pages
    6
  • From page
    233
  • To page
    238
  • Abstract
    On the base of diffusion limited amperometric oxygen sensor (the Clark-type oxygen sensor) a microbial BOD sensor was constructed and the dependence of output signal of device on wastewater biochemical oxygen demand (BOD) was studied. Calibration of BOD sensor was performed in steady-state and dynamic regime of measurements in OECD synthetic wastewater solutions. Results showed broader detection area (up to 230 mg BOD7 L−1) as well as shorter response and regeneration times in dynamic regime of analysis of output signal. The data acquired by steady-state method showed better repeatability, especially in the estimation of BOD7 of real wastewaters. After calibration the BOD sensor was used for quick BOD estimations of municipal wastewaters.
  • Keywords
    BOD biosensor , biochemical oxygen demand , Steady-state response , Dynamic output , Municipal wastewater
  • Journal title
    Sensors and Actuators B: Chemical
  • Serial Year
    2009
  • Journal title
    Sensors and Actuators B: Chemical
  • Record number

    1437659