• Title of article

    Simple and sensitive online detection of triacetone triperoxide explosive

  • Author/Authors

    Lubczyk، نويسنده , , Daniel and Siering، نويسنده , , Carsten and Lِrgen، نويسنده , , Jürgen and Shifrina، نويسنده , , Zinaida B. and Müllen، نويسنده , , Klaus and Waldvogel، نويسنده , , Siegfried R.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2010
  • Pages
    6
  • From page
    561
  • To page
    566
  • Abstract
    Specifically coated high frequency quartz microbalances (200 MHz) were used to investigate a novel sensor system for the detection of TATP (triacetone triperoxide). By combination of differently coated quartz microbalances, it is possible to distinguish TATP in a small test set versus chemically related compounds. The highest discrimination of TATP towards other compounds was achieved with a coating of a phenylene dendrimer. This coating and the high fundamental frequency lead to comparably low limits of detection (1 ppm in a laboratory system) as well as rapid sensor response times.
  • Keywords
    Explosive detection , Quartz microbalance , Triacetone triperoxide , Peroxides , Dendrimers , Supramolecular chemistry
  • Journal title
    Sensors and Actuators B: Chemical
  • Serial Year
    2010
  • Journal title
    Sensors and Actuators B: Chemical
  • Record number