Title of article :
The non-enzymatic determination of glucose using an electrolytically fabricated nickel microparticle modified boron-doped diamond electrode or nickel foil electrode
Author/Authors :
Toghill، نويسنده , , Kathryn E. and Xiao، نويسنده , , Lei and Phillips، نويسنده , , Michael A. and Compton، نويسنده , , Richard G.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2010
Abstract :
A nickel modified boron-doped diamond (Ni-BDD) electrode or nickel foil electrode were used in the non-enzymatic determination of glucose in alkaline solutions. The Ni-BDD electrode was electrodeposited from a 1 mM Ni(NO3)2 solution (pH 5), followed by repeat cycling in KOH. Subsequent analysis utilised the Ni(OH)2/NiOOH redox couple to electrocatalyse the oxidation of glucose.
e was determined to limits of 2.7 μM with a sensitivity of 1.04 μA μM−1 cm−2 at the Ni-BDD electrode. The foil electrode was comparably sensitive achieving a limit of 1.8 μM but a relatively lower sensitivity of 0.67 μA μM−1 cm−2. SEM analysis of the electrodes found the Ni-BDD to be modified by a quasi-random microparticle assembly, with approximately 7.6 μg cm−2 of nickel present on the surface.
Keywords :
Electrocatalysis , Microparticles , nickel , Non-enzymatic , Glucose , Boron-doped diamond
Journal title :
Sensors and Actuators B: Chemical
Journal title :
Sensors and Actuators B: Chemical