• Title of article

    Software-based quantitation of bioassays on CD

  • Author/Authors

    Pallapa Venkataram، نويسنده , , Manu and Ou، نويسنده , , Lily M.L. and Parameswaran، نويسنده , , M. (Ash) and Yu، نويسنده , , Hua-Zhong، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2010
  • Pages
    4
  • From page
    620
  • To page
    623
  • Abstract
    Compact discs (CDs) can be used as rapid, low-cost, high-capacity screening platforms for running bioassays with no modification of the detection hardware (conventional standard optical drive). We describe herein a new protocol to read and quantitate biotin–streptavidin binding assays with a standard optical drive by using a CD-data analysis software (IsoBuster), which identifies erroneous sectors by locating the exact error position bit-by-bit and allows various data formats to be used. The numbers of erroneous sectors increase as a function of the streptavidin concentration in the tested solutions. High spatial accuracy and detection sensitivity (0.3 μg/mL) were achieved.
  • Keywords
    IsoBuster , Compact disc , error detection , Biomarker detection
  • Journal title
    Sensors and Actuators B: Chemical
  • Serial Year
    2010
  • Journal title
    Sensors and Actuators B: Chemical
  • Record number

    1438855