Title of article :
Cross-sensitivity evaluation for ammonia sensing using absorption spectroscopy in the UV region
Author/Authors :
Manap، نويسنده , , H. and Dooly، نويسنده , , G. and O’Keeffe، نويسنده , , S. and Lewis، نويسنده , , E.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2011
Abstract :
This paper describes an optical sensor for the monitoring of ammonia gas. An open path optical technique is used to analyze the absorption lines of ammonia within the ultra-violet region. The optical sensor shows absorption lines comparable to the theoretical ammonia spectra. Cross-sensitivity with carbon dioxide and oxygen gas has been tested and clearly shows that these two gases have no effect on ammonia measurement in the ultra-violet region, between 200 nm and 230 nm. The optical sensor is able to detect ammonia down to 9 ppm with an average error less than 2%. The optical sensor system also shows a fast response time which is less than 4 s.
Keywords :
Optical fibre sensor , Cross-sensitivity , Ammonia sensor
Journal title :
Sensors and Actuators B: Chemical
Journal title :
Sensors and Actuators B: Chemical