Title of article :
An automated microfluidic colourimetric sensor applied in situ to determine nitrite concentration
Author/Authors :
Beaton، نويسنده , , Alexander D. and Sieben، نويسنده , , Vincent J. and Floquet، نويسنده , , Cedric F.A. and Waugh، نويسنده , , Edward M. and Abi Kaed Bey، نويسنده , , Samer and Ogilvie، نويسنده , , Iain R.G. and Mowlem، نويسنده , , Matthew C. and Morgan، نويسنده , , Hywel، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2011
Pages :
6
From page :
1009
To page :
1014
Abstract :
A stand-alone sensor system with integrated sub-systems is demonstrated. The system is portable and capable of in situ reagent-based nutrient analysis. The system is based on a low cost optical detection method, together with an automated microfluidic delivery system that is able to detect nitrite with a limit of detection (LOD) of 15 nM. The sensor was operated in situ at Southampton Dockhead for 57 h (December 2010) and 375 measurements were taken.
Keywords :
Griess nitrite detection , Automated PMMA microfluidic chip , Integrated colourimetric sensor , Spectrophotometry , Environment
Journal title :
Sensors and Actuators B: Chemical
Serial Year :
2011
Journal title :
Sensors and Actuators B: Chemical
Record number :
1439571
Link To Document :
بازگشت