Title of article :
Full-field surface plasmon resonance sensor for high resolution refractive index measurement using common-path heterodyne interferometer
Author/Authors :
Lo، نويسنده , , Y.L. and Chuang، نويسنده , , Ch. R. Lin and S. T. Lin ، نويسنده , , Z.W.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2011
Pages :
5
From page :
400
To page :
404
Abstract :
This study presents a full-field surface plasmon resonance (SPR) sensor induced by attenuated total reflection (ATR)-couple for liquid refractive index measurement. The system adopts a common-path heterodyne interferometer to measure the phase difference between P- and S-wave after passing through the SPR sensor. In order to realize the full-field measurement, it adopts a three-frame integrating-bucket method. The experimental results show great consistency profile between single point and full-field liquid refractive index measurement from 1.330 to 1.340 RIU. It shows that the best sensitivity and resolution of a single pixel in charge couple device (CCD) for liquid refractive measurement are 3.3 × 104 (deg/RIU) and 3.53 × 10−6 (RIU), respectively. As compared with traditional single-point method, the proposed method with a regular CCD has no degradation. Therefore, the system has many applications in chemistry and biology.
Keywords :
surface plasmon resonance , Common-path heterodyne interferometer , Full-field measurements
Journal title :
Sensors and Actuators B: Chemical
Serial Year :
2011
Journal title :
Sensors and Actuators B: Chemical
Record number :
1439997
Link To Document :
بازگشت