• Title of article

    The effect of DNA probe distribution on the reliability of label-free biosensors

  • Author/Authors

    Shinwari، نويسنده , , M.W. and Shinwari، نويسنده , , M.F. and Deen، نويسنده , , M.J. and Selvaganapathy، نويسنده , , P.R.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2011
  • Pages
    7
  • From page
    441
  • To page
    447
  • Abstract
    As the design of label-free DNA biosensors matures, and their sizes reduced to enhance their sensitivity, not much has been researched about the variations in the received signal with the positioning of the probes on the sensitive surface. We approach this issue computationally in this paper. By adopting the finite-element model on a three-dimensional biological field-effect transistor (BioFET) slice, and running Monte-Carlo simulations on the positions of the DNA molecules, we extract the expected variations in the signal. Then, we show that signal-to-noise (SNR) ratio can be low enough to hinder the functionality of the device, placing a limitation on how low the sensitivity of a sensor of a certain size can be.
  • Keywords
    Simulation , Biosensor , Finite-element , DNA , lab-on-chip , label-free
  • Journal title
    Sensors and Actuators B: Chemical
  • Serial Year
    2011
  • Journal title
    Sensors and Actuators B: Chemical
  • Record number