Title of article :
The effect of DNA probe distribution on the reliability of label-free biosensors
Author/Authors :
Shinwari، نويسنده , , M.W. and Shinwari، نويسنده , , M.F. and Deen، نويسنده , , M.J. and Selvaganapathy، نويسنده , , P.R.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2011
Abstract :
As the design of label-free DNA biosensors matures, and their sizes reduced to enhance their sensitivity, not much has been researched about the variations in the received signal with the positioning of the probes on the sensitive surface. We approach this issue computationally in this paper. By adopting the finite-element model on a three-dimensional biological field-effect transistor (BioFET) slice, and running Monte-Carlo simulations on the positions of the DNA molecules, we extract the expected variations in the signal. Then, we show that signal-to-noise (SNR) ratio can be low enough to hinder the functionality of the device, placing a limitation on how low the sensitivity of a sensor of a certain size can be.
Keywords :
Simulation , Biosensor , Finite-element , DNA , lab-on-chip , label-free
Journal title :
Sensors and Actuators B: Chemical
Journal title :