• Title of article

    Prothrombin time tests on a microfluidic disc analyzer

  • Author/Authors

    Shih، نويسنده , , Chih-Hsin and Lu، نويسنده , , Chien-Hsing and Wu، نويسنده , , Jia-Huei and Lin، نويسنده , , Chia-Hui and Wang، نويسنده , , Jiunn-Min and Lin، نويسنده , , Chi-Yu، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2012
  • Pages
    7
  • From page
    1184
  • To page
    1190
  • Abstract
    Prothrombin time test is used to measure the time required for blood to coagulate and is essential for monitoring appropriate drug dosing in patients taking oral anti-coagulants. To reduce the costs of reagents and improve the stability of tests, a microfluidic disc analyzer (MDA) has been developed. This analyzer can process small volumes of liquid samples (less than 10 μl total volume required) and performs effective and fast mixing (within one second) under centrifugation. During the mixing process, two reactants can also maintain a constant ratio of flowrate. One hundred and one clinical samples were tested by the MDA and the Sysmex CA1500 coagulation analyzer, which was frequently used as a reference machine, for comparison. The test results from these two instruments show good correlation and agreement.
  • Keywords
    Microfluidic disc analyzer , Prothrombin time test , Mixing , Centrifugation , International normalized ratio
  • Journal title
    Sensors and Actuators B: Chemical
  • Serial Year
    2012
  • Journal title
    Sensors and Actuators B: Chemical
  • Record number