Title of article
Detection limits of chemical sensors: Applications and misapplications
Author/Authors
Loock، نويسنده , , Hans-Peter and Wentzell، نويسنده , , Peter D.، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2012
Pages
7
From page
157
To page
163
Abstract
The limit of detection (LOD) and the sensitivity of a chemical sensor are defined using IUPAC guidelines. The LOD from simulated and experimental data is calculated from a calibration curve using a simple statistical model that was implemented into a spreadsheet program. This definition of the LOD is compared with the commonly used definition of the LOD, which is based on the product of sensitivity and the theoretical instrument resolution.
Keywords
Sensitivity , Fiber optic sensors , confidence interval , Photoacoustic sensor , Calibration Curve , Absorption sensor , Limit of detection , Refractive index sensor
Journal title
Sensors and Actuators B: Chemical
Serial Year
2012
Journal title
Sensors and Actuators B: Chemical
Record number
1440982
Link To Document