Title of article :
Electronic noses calibration procedure in the context of a multicentre medical study
Author/Authors :
Santonico، نويسنده , , M. and Pennazza، نويسنده , , G. and Capuano، نويسنده , , R. and Falconi، نويسنده , , C. J. Vink، نويسنده , , T.J. and Knobel، نويسنده , , H.H. and Van der Schee، نويسنده , , M.P. and Sterk، نويسنده , , P.J. and Montuschi، نويسنده , , P. and D’Amico، نويسنده , , A.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2012
Abstract :
An overview of the scientific literature in the medical field tells us that a large part of electronic nose applications is devoted to breath analysis. A network based study can help testing the validity of this strategy when using many different devices based on identical or different technologies, in view of a use in real clinical practice.
rst step is the instrument calibration using a set of key-compounds. In this work a gas sensor array based on quartz microbalance transducers functionalized with metalloporphyrins (ROTV e-nose), and a Cyranose is used simultaneously in a calibration experiment with three ad hoc selected compounds: ethanol, hexane and ethyl acetate, at concentration levels around 1 ppm. These tests have demonstrated that limits of detection down to tens of ppb are possible. Moreover, a mapping between the two instruments has been performed through the calculation of a model based on Cyranose data, and applied to the ROTV e-nose data, for the prediction of compound concentrations. This test has shown a good ability in concentrations prediction, with an error lower than 10 ppb.
clusion, this study provides the first step toward quality assurance of e-nose data in the medical setting.
Keywords :
Medical applications , Limit of detection , Permeation tubes , Electronic nose , volatile compounds