Title of article :
High speed and high resolution chemical imaging based on a new type of OLED-LAPS set-up
Author/Authors :
Werner، نويسنده , , Carl Frederik and Wagner، نويسنده , , Torsten and Miyamoto، نويسنده , , Ko-ichiro and Yoshinobu، نويسنده , , Tatsuo and Schِning، نويسنده , , Michael J.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2012
Pages :
5
From page :
118
To page :
122
Abstract :
Light-addressable potentiometric sensors (LAPS) are field-effect-based sensors. A modulated light source is used to define the particular measurement spot to perform spatially resolved measurements of chemical species and to generate chemical images. In this work, an organic-LED (OLED) display has been chosen as a light source. This allows high measurement resolution and miniaturisation of the system. A new developed driving method for the OLED display optimised for LAPS-based measurements is demonstrated. The new method enables to define modulation frequencies between 1 kHz and 16 kHz and hence, reduces the measurement time of a chemical image by a factor of 40 compared to the traditional addressing of an OLED display.
Keywords :
OLED display , Light-addressable potentiometric sensor (LAPS) , Chemical sensor , Chemical imaging