Title of article
The sensing mechanism and detection of low concentration acetone using chitosan-based sensors
Author/Authors
Nasution، نويسنده , , Tulus Ikhsan and Nainggolan، نويسنده , , Irwana and Hutagalung، نويسنده , , Sabar Derita and Ahmad، نويسنده , , Khairel Rafezi and Ahmad، نويسنده , , Zainal Arifin، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2013
Pages
7
From page
522
To page
528
Abstract
This paper aims to discuss the usage of chitosan film sensors to detect acetone concentrations in human breath, in order to precisely diagnose diabetes mellitus in patients. Acetone concentration in the breath varies from 0.3 to 0.9 ppm in healthy people to more than 1.8 ppm for diabetics. This makes acetone a suitable chemical marker for diabetes diagnosis. Therefore, the preliminary study on the electrical laboratory testing of the chitosan film sensor properties to acetone vapor-contaminated air in range of 0.1–100 ppm was carried out at room temperature (∼25–30 °C) in normal air. Our results suggested that the proposed acetone-based gas sensor can operate at room temperature with a high performance demonstrated by good response, recovery, stability and repeatability. This trouble free, painless and steadfast technique will improve the current gold standard in diagnosing diabetes, enabling quick and early detection.
Keywords
Acetone vapor , Electrical properties , diabetes mellitus , electrochemical deposition , Chitosan-based acetone sensor
Journal title
Sensors and Actuators B: Chemical
Serial Year
2013
Journal title
Sensors and Actuators B: Chemical
Record number
1441457