Title of article :
Self-made non-enzymatic silver electrode from recordable CDs for fast detection of glucose in blood
Author/Authors :
Chen، نويسنده , , Yuxing and Liu، نويسنده , , Lin and Wang، نويسنده , , Meirong and Wang، نويسنده , , Chengyin and Hu، نويسنده , , Xiaoya and Wang، نويسنده , , Guoxiu، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2013
Pages :
7
From page :
555
To page :
561
Abstract :
An electrochemical sensor based on a self-made electrode from recordable CDs was developed for the non-enzymatic detection of glucose by chronoamperometry. We discussed the amperometric current response of glucose with the change of potential, pH and electrode area, and determined the optimum detection conditions. The current response measurements were performed in a phosphate buffered solution (pH 6.5) with a potential of −0.50 V, and presented a linearity over the range of 0.5–13 mmol/L (r = 0.996). The experimental results of the designed sensor demonstrate that this method has merits such as simple operation, low cost and rapid responses. The results of detecting glucose in blood samples were satisfactory.
Keywords :
Recordable CDs , chronoamperometry , Non-enzymatic sensor , Blood , Glucose
Journal title :
Sensors and Actuators B: Chemical
Serial Year :
2013
Journal title :
Sensors and Actuators B: Chemical
Record number :
1441461