Title of article
Development of field-deployable QCL sensor for simultaneous detection of ambient N2O and CO
Author/Authors
Li، نويسنده , , Jingsong and Parchatka، نويسنده , , Uwe and Fischer، نويسنده , , Horst، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2013
Pages
9
From page
659
To page
667
Abstract
We report the development of a field deployable quantum cascade laser (QCL) sensor for simultaneous high sensitivity and high precision detection of the atmospheric trace gases nitrous oxide (N2O) and carbon monoxide (CO). The system relies on a continuous-wave room temperature quantum-cascade laser source emitting at a wavelength of around 4.56 μm, where strong fundamental absorption bands occur for both species. The detection technique is based on second harmonic wavelength modulation spectroscopy and long multi-pass absorption technique. Primary laboratory tests over a 5-day period have been performed to estimate the achievable detection limits and the signal reproducibility levels in view of possible applications for in situ and real time atmospheric measurements. The short-term precision of 1.29 and 1.98 ppbv at 1 Hz under laboratory conditions was demonstrated, corresponding to line centre absorbance of 1.48 × 10−4 and 1.80 × 10−4 for CO and N2O, respectively.
Keywords
QCL , nitrous oxide , CARBON MONOXIDE , Wavelength modulation spectroscopy