Title of article :
Effect of the geometry of a working electrode on the behavior of a planar amperometric NO2 sensor based on solid polymer electrolyte
Author/Authors :
Kubersk?، نويسنده , , P. and Ham??ek، نويسنده , , A. and Ne?p?rek، نويسنده , , S. and Soukup، نويسنده , , R. and Vik، نويسنده , , R.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2013
Abstract :
An amperometric NO2 sensor with a new type of solid polymer electrolyte (SPE) and a glassy carbon working electrode is presented. The electrolyte is based on 1-ethyl-3-methylimidazolium bis(trifluoromethylsulfonyl)imide [C2mim][NTf2] ionic liquid immobilized in poly(vinylidene fluoride) matrix [PVDF]. The analyte, gaseous nitrogen dioxide, was detected by reduction at −500 mV vs. a platinum pseudoreference electrode. The sensor exhibited a linear behavior in the whole tested range, i.e., 0–10 ppm, and its sensitivity was 0.66 × 10−6 A/ppm. The rise/recovery times were of the order of tens of seconds. In addition, the influence of the geometry (area and thickness) of the carbon working electrode on sensor response during short-term and long-term exposure to nitrogen dioxide was investigated. The area of the working electrode strongly influenced the stability and noise of the sensor: the smaller the electrode area, the better the stability of the sensor and the lower the noise level. The response times were influenced by the thickness of the carbon working electrode: the thicker the electrode, the longer the response time. It was also demonstrated that glassy carbon is a good candidate for the preparation of the working electrode. The use of a platinum pseudoreference electrode simplifies sensor fabrication and does not have any negative effect on sensor stability.
Keywords :
Ionic liquid , Amperometric sensor , nitrogen dioxide , Solid polymer electrolyte
Journal title :
Sensors and Actuators B: Chemical
Journal title :
Sensors and Actuators B: Chemical