• Title of article

    Acetylcholinesterase-based conductometric biosensor for determination of aflatoxin B1

  • Author/Authors

    Soldatkin، نويسنده , , O.O. and Burdak، نويسنده , , O.S. and Sergeyeva، نويسنده , , T.A. and Arkhypova، نويسنده , , V.M. and Dzyadevych، نويسنده , , S.V. and Soldatkin، نويسنده , , A.P.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2013
  • Pages
    5
  • From page
    999
  • To page
    1003
  • Abstract
    The theoretical and experimental studies were carried out, which resulted in the development of acetylcholinesterase-based conductometric biosensor for determination of aflatoxin B1. We established a linear range of substrate measurement (0.25–1 mM) and optimal concentration of acetylcholine for inhibitory analysis (1 mM). The sensitivity of developed biosensor to aflatoxin B1 was tested, the calibration curve was plotted for the dependence of residual activity of the acetylcholinesterase-based bioselective element on the toxin concentration. Detection limit of aflatoxin B1 was 0.05 μg/ml. It was shown that the developed biosensor was characterized by high signal reproducibility of direct determination of the main substrate as well as of the inhibitory analysis of aflatoxin B1 (the error of detection was no more than 7%).
  • Keywords
    Inhibitory analysis , Environmental monitoring , Aflatoxin B1 , Acetylcholinesterase , Biosensor , Conductometric transducer
  • Journal title
    Sensors and Actuators B: Chemical
  • Serial Year
    2013
  • Journal title
    Sensors and Actuators B: Chemical
  • Record number

    1443082