Title of article
Electroanalysis of lincomycin using boron-doped diamond thin film electrode applied to flow injection system
Author/Authors
Boonsong، نويسنده , , Kanokporn and Chuanuwatanakul، نويسنده , , Suchada and Wangfuengkanagul، نويسنده , , Nattakarn and Chailapakul، نويسنده , , Orawon Chailapakul، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2005
Pages
6
From page
627
To page
632
Abstract
The electroanalysis of lincomycin was investigated using boron-doped diamond thin film (BDD) electrodes. First, the electrochemistry of lincomycin was studied by cyclic voltammetry as a function of pH of the solution, scan rate and the concentration of lincomycin. Comparison experiments were carried out using glassy carbon electrode. Boron-doped diamond thin film electrode provided well-resolved oxidation irreversible cyclic voltammogram. It was found that the peak position was 1.2 V (versus Ag/AgCl reference electrode). Second, the amperometric detection with BDD electrode was coupled with the flow injection analysis for the determination of lincomycin. The linear range of 0.5–125 μM and the detection limit of 0.02 μM were obtained. The percent recovery was shown in a range of 96–103. This purposed method was also applied to a drug formulation sample. It was found that the results (293 mg mL−1) were comparable to those labeled (300 mg mL−1).
Keywords
Boron-doped diamond thin film electrode , Cyclic voltammetry , lincomycin , Flow injection with amperometric detection
Journal title
Sensors and Actuators B: Chemical
Serial Year
2005
Journal title
Sensors and Actuators B: Chemical
Record number
1443374
Link To Document