    The calibration temperature of pH-glass electrodes: significance for meat quality classification

  • Author/Authors

    Karlsson، نويسنده , , Anders H. and Rosenvold، نويسنده , , Katja، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2002
  • Pages
    5
  • From page
    497
  • To page
    501
  • Abstract
    In this short communication we show important factors concerning pH measurements in pork. The purpose of the study was to investigate the effect of the calibration temperature of pH-buffer and of the pH-electrode on pH measured 1–120 min post mortem in pork, and to evaluate its significance for prediction of meat quality. The calibration temperature of pH-electrodes as well as the sample temperature is essential for pH measurements, as pH measurements are highly temperature-dependent. This is especially important when using pH measurement for prediction of meat quality or when different experiments are compared. We have used two calibration temperatures of 15 °C (normal) and 35 °C (modified) in the present investigation. Based on our findings, we therefore recommend a buffer and electrode temperature of 35 °C as a standard procedure for measuring pH early post mortem. As a minimum the pH should be presented along with the measuring and calibrating temperatures.
  • Keywords
    Classification , meat quality , pig , PH , Calibration