• Title of article

    Phenols monitoring and Hill coefficient evaluation using tyrosinase-based amperometric biosensors

  • Author/Authors

    Stanca، نويسنده , , Sarmiza Elena and Popescu، نويسنده , , Ionel Catalin Popescu، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2004
  • Pages
    6
  • From page
    47
  • To page
    52
  • Abstract
    Sensitive amperometric biosensors for phenols compounds, based on tyrosinase (polyphenoloxidase, PPO) immobilized on a Pt electrode in an electropolymerized poly-amphiphilic pyrrole matrix or cross-linked with glutaraldehyde, were constructed and compared. Steady-state amperometric measurements, performed at −50 mV vs. SCE in aqueous phosphate buffer containing LiClO4 0.1 M (pH 7) as well as in a chloroform solution containing 0.1 M C6H5CH2N(CH3)3Cl, were used in order to compare the electroanalytical and kinetic parameters of the investigated amperometric biosensors in aqueous and nonaqueous media. It was established that the polypyrrole matrix has a higher efficiency for enzyme retention resulting in higher bioelectrode sensitivity, both in aqueous buffer (690 μA M−1) and in chloroform (149 μA M−1).
  • Keywords
    tyrosinase , Amperometric biosensors , Phenols
  • Journal title
    Bioelectrochemistry
  • Serial Year
    2004
  • Journal title
    Bioelectrochemistry
  • Record number

    1451204