Title of article :
Changes of the solution pH due to exposure by high-voltage electric pulses
Author/Authors :
Alexandrina S. Saulis، نويسنده , , Gintautas and Lap?، نويسنده , , Remigijus and Pranevi?i?t?، نويسنده , , Rita and Mickevi?ius، نويسنده , , Donatas، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2005
Abstract :
The change of the pH of a NaCl solution (139–149 mM NaCl) buffered with 5–15 mM sodium phosphates (pH 7.4) during electromanipulation was studied. It has been determined that an increase in the pH value of electroporation solution of a whole chamber volume, caused by the application of electric field pulses, commonly used in cell electromanipulation procedures, can exceed 1–2 pH units.
l materials for the cathode were tested. In all cases a stainless steel anode was utilized. The aluminum cathode gave a two-fold greater ΔpH in comparison with platinum, copper or stainless steel cathodes. In addition, a substantial release of aluminum (up to 1 mg/l) from the cathode was observed. It has also been found that the shift in pH depended on the medium conductivity: ΔpH of the solution, in which sucrose was substituted for NaCl, was about 5 times less.
basis of the results obtained here, to avoid the plausible undesirable consequences of the cathodic electrolysis processes, in particular under the conditions of strong electric treatment, it could be recommended that chambers with aluminum electrodes not be utilized and one should use strongly buffered solutions of low conductivity and alternating current (sine or square wave) bipolar electric pulses.
Keywords :
Cell electromanipulation , Electroporation , pH shift , electrodes , Electrolysis reactions