    Development of cholesterol biosensor based on immobilized cholesterol esterase and cholesterol oxidase on oxygen electrode for the determination of total cholesterol in food samples

  • Author/Authors

    Basu، نويسنده , , Anjan Kumar and Chattopadhyay، نويسنده , , Parimal and Roychoudhuri، نويسنده , , Utpal and Chakraborty، نويسنده , , Runu، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2007
  • Pages
    5
  • From page
    375
  • To page
    379
  • Abstract
    The development of a cholesterol biosensor by co-immobilization of cholesterol esterase (ChEt) and cholesterol oxidase (ChOX) on oxygen electrode is described. The electrode consists of gold cathode and Ag/AgCl anode. The enzymes were immobilized by cross-linking with glutaraldehyde and Bovine Serum Albumin (BSA). The immobilized enzymatic membrane was attached to the tip of the electrode by a push cap system. The optimum pH and temperature of the sensor was determined, these are 6 and 25 °C respectively. The developed sensor was calibrated from 1–75 mg/dl of cholesterol palmiate and found linear in the range of 2–50 mg/dL. The calibration curve was drawn with Vi (ppm/min)(initial velocity) vs different concentrations of cholesterol palmiate (mg/dL). The application of the sensor to determine the total cholesterol in different real food samples such as egg, meat was investigated. The immobilized enzymatic layer can be reused over 30 times and the stability of the enzymatic layer was studied up to 9 weeks.
  • Keywords
    ChEt , Cholesterol biosensor , Food sample analysis , ChOX , stability
  • Journal title
    Bioelectrochemistry
  • Serial Year
    2007
  • Journal title
    Bioelectrochemistry
  • Record number