Title of article
Monitoring of glucose and glutamate using enzyme microstructures and scanning electrochemical microscopy
Author/Authors
Mure?an، نويسنده , , Laura and Nistor، نويسنده , , Mihaela and G?sp?r، نويسنده , , Szilveszter and Popescu، نويسنده , , Ionel C?t?lin and Cs?regi، نويسنده , , Elisabeth، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2009
Pages
6
From page
81
To page
86
Abstract
Glucose oxidase and glutamate oxidase lines, with typical width of 100 µm, were patterned on gold surfaces using a micro-dispensing system, by shooting 100 pl droplets of the corresponding enzyme solutions. The probe of a scanning electrochemical microscope (SECM) was then carefully positioned in the close proximity of the enzyme microstructure and poised to + 600 mV vs. Ag/AgCl, KCl 0.1 M. The H2O2, generated by the enzyme lines at different concentrations of glucose and glutamate in the surrounding solution, was sequentially monitored. Reproducible calibration curves for glucose and glutamate were obtained in one single experiment, proving that the combination of enzyme microstructures with SECM can provide a new way of achieving multianalyte detection.
Keywords
Glucose oxidase , Glutamate oxidase , H2O2 , Enzyme microstructures , Scanning electrochemical microscopy