Title of article
Assessment of sirolimus concentrations in whole blood by high-performance liquid chromatography with ultraviolet detection
Author/Authors
Cattaneo، نويسنده , , Dario and Perico، نويسنده , , Norberto and Gaspari، نويسنده , , Flavio، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2002
Pages
8
From page
187
To page
194
Abstract
A novel, reversed-phase high-performance liquid chromatographic (HPLC) method is described for the analysis of sirolimus (SRL) in whole blood. The samples were purified by precipitating blood matrix with zinc sulfate, SRL was then extracted with acetone followed by solid-phase extraction. The method was linear over a range of 1–100 ng/ml and the lower limit of quantification was 2.5 ng/ml. The coefficient of variation (within day) was below 8.0% for the lowest SRL concentration. The day-to-day coefficient of variation was below 6.6%. The assay did not show interference peaks with immunosuppressive drugs commonly given to transplant patients. With the simplified extraction procedure described, 60 samples (including controls and calibration curve) can be quantified in a day. The sensitivity and rapidity of this analytical procedure makes it useful for routine therapeutic monitoring of SRL.
Keywords
SIROLIMUS
Journal title
Journal of Chromatography B
Serial Year
2002
Journal title
Journal of Chromatography B
Record number
1453805
Link To Document