• Title of article

    High-performance liquid chromatography coupled with negative ion tandem mass spectrometry for determination of pravastatin in human plasma

  • Author/Authors

    Zhu، نويسنده , , Zhimeng and Neirinck، نويسنده , , Len، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2003
  • Pages
    8
  • From page
    133
  • To page
    140
  • Abstract
    A new method, using high-performance liquid chromatography/ion electrospray (negative ion) mass spectrometry, has been developed for the determination of a hydrophilic liver-specific inhibitor of the enzyme 3-hydroxy-3-methylglutaryl coenzyme A reductase, pravastatin in human plasma. In this method, plasma samples were prepared by a solid-phase extraction on C18 Bond Elut cartridge. Chromatography was carried out with a Zorbax C8 column. Simple isocratic chromatography conditions were used. The method has been validated in a linear range of 0.25–300 ng/ml with a coefficient of variation of 0.6–3.4%. The overall recovery was 90.5% for pravastatin and 90.8% for the internal standard β-hydroxy-lovastatin. The method is simple and reliable with a total run time of less than 2 min.
  • Keywords
    Pravastatin
  • Journal title
    Journal of Chromatography B
  • Serial Year
    2003
  • Journal title
    Journal of Chromatography B
  • Record number