Title of article :
Development of a liquid chromatography–mass spectrometry method for monitoring the angiotensin-converting enzyme inhibitor lisinopril in serum
Author/Authors :
Tsakalof، نويسنده , , Andreas and Bairachtari، نويسنده , , Kyriaki and Georgarakis، نويسنده , , Manolis، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2003
Pages :
8
From page :
425
To page :
432
Abstract :
In this study, a sensitive, specific, precise and accurate method for lisonopril quantitative determination in human serum was developed and validated. The method comprises lisinopril isolation from serum by means of solid-phase extraction followed by its quantification by liquid chromatography–mass spectrometry. Chromatographic separation was performed at 55 °C on Kromasil C18 5 μm 250×3.2 mm HPLC column with mobile phase composed of 50 mM ammonium formate buffer (pH 3)–acetonirile–methanol (72:7:21, v/v/v). A Finnigan AQA benchtop mass spectrometer with a pneumatically assisted electrospray (ES) interface and a single quadrupole mass filter was used to detect and quantify lisinopril in column effluent. Ion signals were acquired by selected ion monitoring of the protonated lisinopril ion m/z=406.5 (M+1). The detector response was linear with r>0.9993 in the investigated concentration range 6–150 ng/ml. The mean recovery of lisinopril from serum samples was 88%. The limit of quantitation for lisinopril was 6 ng/ml with a signal-to-noise ratio at this concentration level S/N=34.75±3.9 (n=4).
Keywords :
Lisinopril
Journal title :
Journal of Chromatography B
Serial Year :
2003
Journal title :
Journal of Chromatography B
Record number :
1454672